FDA · Medical device recall
GE Healthcare Definium 8000 Digital Radiographic System…
- Date posted by the FDA
- Sept. 25, 2008
- Status
- Terminated
- Filed by
- GE Healthcare
- Product
- GE Healthcare Definium 8000 Digital Radiographic System. Formerly GE Healthcare Revolution XR/d with Tomosynthesis.
- Reason
- Loss of image data: Software anomaly in the Processing Software on the Definium 8000 system may impact patient safety when using the VolumeRAD advanced application (option). The slice visualization of VolumeRAD exams acquired on the Wallstand receptor will be offset by 17mm. This prevents the visualization of anatomy WITHIN a distance of 17mm from the wallstand patient barrier (receptor cover).
- Distribution pattern
- Worldwide distribution: USA, Australia, Belgium, Canada, China, Finland, France, Germany, Great Britain, Italy, Netherlands, Spain, and Sweden.
- Quantity
- 127
- Codes and lots
- 00000002047M39, 00000005487M34, 00000096986DP%, 00000100740WK6, 00000TA41882-7, 00000TA41972-3, 00000TA41984-4, 00000TA41984-6, 00000TA42274-8, 00000TA42869-6, 00001004330WK5, 00001004332WK1, 00001004504WK5, 00001004505WK2, 00001004510WK2, 00001004513WK6, 00001004764WK5, 00001004765WK2, 00001004766WK0, 00001004768WK6, 00001004770WK2, 00001004772WK8, 00001005020WK1, 00001005022WK7, 00001005024WK3, 00001005030WK0, 00001005031WK8, 00001005035WK9, 00001005181WK1, 00001005182WK9, 00001005188WK6, 00001005260WK3, 00001005339WK5, 00001005758WK6, 00001005914WK5, 00001006042WK4, 00001006474WK9, 00001006635WK5, 00001006637WK1, 00001006638WK9, 00001006668WK6, 00001006871WK6, 00001007161WK1, 00001007166WK0, 00001007403WK7, 00001007404WK5, 00001007480WK5, 00001007536WK4, 00001007644WK6, 00001007650WK3, 00001007907WK7, 00001008312WK9, 00001008314WK5, 00001008497WK8, 00001008684WK1, 00001008807WK8, 00001008813WK6, 00001009140WK3, 00001009143WK7, 00001009145WK2, 00001009216WK1, 00001009268WK2, 00001009375WK5, 00001009380WK5, 00001009382WK1, 00001009709WK5, 00001009818WK4, 00001009875WK4, 00001009934WK9, 00001009976WK0, 00001010085WK7, 00001010099WK8, 00001010161WK6, 00001010239WK0, 00001010294WK5, 00001010295WK2, 00001010297WK8, 00001010431WK3, 00001010488WK3, 00001010524WK5, 00001010628WK4, 00001010629WK2, 00001010687WK0, 00001010693WK8, 00001010695WK3, 00001010906WK4, 00001010907WK2, 00001010964WK3, 00001011084WK9, 00001011085WK6, 00001011092WK2, 00001011215WK9, 00001011285WK2, 00001011289WK4, 00001011386WK8, 00001011455WK1, 00001011488WK2, 00001011490WK8, 00001011647WK3, 00001011654WK9, 00001011700WK0, 00001011712WK5, 00001011822WK2, 00001011825WK5, 00001011827WK1, 00001011828WK9, 00001011829WK7, 00001011832WK1, 00001012012WK9, 00001012119WK2, 00001012286WK9, 00001012449WK3, 00001012450WK1, 00001012453WK5, 00001012923WK7, 00001013064WK9, 00001013067WK2, 00001013100WK1, 00001013101WK9, 00001013134WK0, 00001013136WK5, 00001013223WK1, 00001013334WK6, 00001013336WK1, 00001013622WK4, 00001014310WK5, 00001014936WK7, 00001016147WK9
- Recall started
- March 7, 2008
- Device type
- System, X-Ray, Stationary
- Root cause (FDA)
- Software design
- What the company told customers
- GE sent a Product Safety Notification, dated March 7, 2008, to all customers. The letter described the safety issue, products affected, and stated that a GE Healthcare Field Engineer will schedule a visit to each facility to install a software modification.
- Recall number
- Z-1822-2008
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-1822-2008. The agency record governs; check it and the company before acting.