Menu
FDA · Medical device recall

GE Healthcare Centricity Universal Viewer Zero Footprint…

Nov. 13, 2024 Class II

Date
Nov. 13, 2024
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing
Filed by
GE Healthcare
Product
GE Healthcare Centricity Universal Viewer Zero Footprint, Model Numbers: 1) 2089507-083, GDZH-JINWAN-Centricity 2) 2089507-093, 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02268286, 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02030900, 1495-1-Centricity Universal Viewer Zero Footprint / ZFP-02347996, 423834-Centricity Universal Viewer Zero Footprint / ZFP-02286658, 1407636-Centricity Universal Viewer Zero Footprint / ZFP-02304121, 651240-Centricity Universal Viewer Zero Footprint / ZFP-02173181, 268034-Centricity Universal Viewer Zero Footprint / ZFP-02402520; 3) 2089507-098, JP-310027-Centricity Universal Viewer Zero Footprint / ZFP-02033125; 4) 2089507-109, 3909-2-Centricity Universal Viewer Zero Footprint / ZFP-20095312, 3909-2-Centricity Universal Viewer Zero Footprint / ZFP-20095313, 327750-Centricity Universal Viewer Zero Footprint / ZFP-02184508; 5) 2089507-111, 105468-1-Centricity Universal Viewer Zero Footprint / ZFP-02254840, 1495-3-Centricity Universal Viewer Zero Footprint / ZFP-00096090, JP-340111-Centricity Universal Viewer Zero Footprint / ZFP-01275379, JP-275353-Centricity Universal Viewer Zero Footprint / ZFP-02042636; 6) 2089507-112, 3922-25-Centricity Universal Viewer Zero Footprint / ZFP-01290685, 3922-17-Centricity Universal Viewer Zero Footprint / ZFP-00098229, 1503586-Centricity Universal Viewer Zero Footprint / ZFP-21016603, RU0994-Centricity Universal Viewer Zero Footprint / ZFP-02331255; 7) 2089507-114, 1461-1-Centricity Universal Viewer Zero Footprint / ZFP-02360229, 1461-1-Centricity Universal Viewer Zero Footprint / ZFP-01803680, 3728-2-Centricity Universal Viewer Zero Footprint / ZFP-01225897, 3728-2-Centricity Universal Viewer Zero Footprint / ZFP-01225898, 3728-2-Centricity Universal Viewer Zero Footprint / ZFP-01225904, 3728-2-Centricity Universal Viewer Zero Footprint / ZFP-01371500, M51725-Centricity Universal Viewer Zero Footprint / ZFP-01300873, M1781854-Centricity Universal Viewer Zero Footprint / ZFP-01553772, 609334-Centricity Universal Viewer Zero Footprint / ZFP-01282257, UJJF01-Centricity Universal Viewer Zero Footprint / ZFP-22186138, CMC-Centricity Universal Viewer Zero Footprint / ZFP-01820241, JP-100014-Centricity Universal Viewer Zero Footprint / ZFP-01890709, KW1102-Centricity Universal Viewer Zero Footprint / ZFP-02191178, 59219_ERPerror-Centricity Universal Viewer Zero Footprint / ZFP-02327073, QA1114-Centricity Universal Viewer Zero Footprint / ZFP-02306397, SAU1433-Centricity Universal Viewer Zero Footprint / ZFP-01457278, TTSH-CVIS-Centricity Universal Viewer Zero Footprint / ZFP-02041523, NE4651-Centricity Universal Viewer Zero Footprint / ZFP-00038599, NE6703-Centricity Universal Viewer Zero Footprint / ZFP-00038552, NE6731-Centricity Universal Viewer Zero Footprint / ZFP-00038461, NE3006-Centricity Universal Viewer Zero Footprint / ZFP-00038362, NE2196-Centricity Universal Viewer Zero Footprint / ZFP-00038588, NPV738-Centricity Universal Viewer Zero Footprint / ZFP-00038637, NE2031-Centricity Universal Viewer Zero Footprint / ZFP-00038578, NE7405-Centricity Universal Viewer Zero Footprint / ZFP-00038372, NE2162-Centricity Universal Viewer Zero Footprint / ZFP-00038563, NE4878-
Reason
GE HealthCare has become aware of an issue in Centricity Universal Viewer Zero Footprint Client (ZFP) where there is a potential security vulnerability which could allow a malicious actor to access the system and potentially manipulate patient data.
Distribution pattern
Worldwide distribution - United States Nationwide and the countries of Algeria, Argentina, Brazil, China, Dominican Republic, France, Germany, India, Indonesia, Ireland, Israel, Japan, Kuwait, Malaysia, Mexico, Panama, Qatar, Russia, Saudi Arabia, Singapore, Spain, United Kingdom.
Quantity
120 units
Codes and lots
UDI/DI 00840682102988 for all models: 1) 2089507-083, Installed Product ID Numbers: GDZH-JINWAN-Centricity Universal Viewer Zero Footprint / ZFP-02171477, 103378-Centricity Universal Viewer Zero Footprint / ZFP-20039058, 1664823-Centricity Universal Viewer Zero Footprint / ZFP-20080654, 493356-Centricity Universal Viewer Zero Footprint / ZFP-20085691, Medireport-Centricity Universal Viewer Zero Footprint / ZFP-02291850, Medireport-Centricity Universal Viewer Zero Footprint / ZFP-02291877, Medireport-Centricity Universal Viewer Zero Footprint / ZFP-02292580, 81077-Centricity Universal Viewer Zero Footprint / ZFP-20680055, IL1469-Centricity Universal Viewer Zero Footprint / ZFP-02156638, JP-110145-Centricity Universal Viewer Zero Footprint / ZFP-01275393, 1333629-Centricity Universal Viewer Zero Footprint / ZFP-01300332; 2) 2089507-093, Installed Product ID Numbers: 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02268286, 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02030900, 1495-1-Centricity Universal Viewer Zero Footprint / ZFP-02347996, 423834-Centricity Universal Viewer Zero Footprint / ZFP-02286658, 1407636-Centricity Universal Viewer Zero Footprint / ZFP-02304121, 651240-Centricity Universal Viewer Zero Footprint / ZFP-02173181, 268034-Centricity Universal Viewer Zero Footprint / ZFP-02402520; 3) 2089507-098, Installed Product ID Numbers: JP-310027-Centricity Universal Viewer Zero Footprint / ZFP-02033125; 4) 2089507-109, Installed Product ID Numbers: 3909-2-Centricity Universal Viewer Zero Footprint / ZFP-20095312, 3909-2-Centricity Universal Viewer Zero Footprint / ZFP-20095313, 327750-Centricity Universal Viewer Zero Footprint / ZFP-02184508; 5) 2089507-111, Installed Product ID Numbers: 6) 2089507-112, Installed Product ID Numbers: 7) 2089507-114, Installed Product ID Numbers: 8) 2089507-115, Installed Product ID Numbers: 9) 2089507-116, Installed Product ID Numbers: 10) 2089507-117, Installed Product ID Numbers: 11) 2089507-118, Installed Product ID Numbers: 12) 2089507-119, Installed Product ID Numbers:
Recall started
Oct. 3, 2024
Device type
System, Image Processing, Radiological
Root cause (FDA)
Software Manufacturing/Software Deployment
What the company told customers
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to it consignees on 10/03/2024 via letter issued through a traceable means. The notice explained the issue and provided the following actions be to be implemented until the devices can be corrected: Option 1: Restrict access to these systems from the internet. or Option 2: Implement additional safeguards such as a VPN. For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or a local Service Representative. GE HealthCare will correct all affected products at no cost. A GE HealthCare representative will make contact to arrange for the correction.
Recall number
Z-0318-2025

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-0318-2025. The agency record governs; check it and the company before acting.

Report a mistake on this page