FDA · Medical device recall
CardioLab/ComboLab Recording Systems
- Date
- Nov. 6, 2019
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Terminated
- Filed by
- GE Healthcare, LLC
- Product
- CardioLab/ComboLab Recording Systems
- Reason
- Potential for failure of the patient leakage current test. There is a potential that if another device with electrical connectivity (e.g., ablation device, ECG monitor) fails, then the CLab II Plus Amplifier will not prevent an electric current from completing a circuit, which could cause an electrical shock to a patient.
- Distribution pattern
- Nationwide distribution to CA, CT, FL, IN, MD, MI , MN, MO, NC, NE, NJ, OH, PA, TX, WI. International distribution to Canada, France, India, Japan, Korea (Republic Of), Mexico, Spain, Taiwan.
- Quantity
- 43 (27 US, 16 OUS)
- Codes and lots
- Model Number: 2088700-002, No UDI Number, Lot Numbers: RXJ19122006TA, RXJ19122002TA, RXJ19122010TA, RXJ19122001TA, RXJ19122005TA, RXJ19122007TA; Model Number: 2088700-004, UDI Number (Lot Numbers): 01008406821196031119050021RXJ19204004TA (RXJ19204004TA), 01008406821196031119040021RXJ19154006TA (RXJ19154006TA), 01008406821196031119050021RXJ19224011TA (RXJ19224011TA), 01008406821196031119050021RXJ19224017TA (RXJ19224017TA), 01008406821196031119050021RXJ19224014TA (RXJ19224014TA), 01008406821196031119040021RXJ19154009TA (RXJ19154009TA), 01008406821196031119040021RXJ19154011TA (RXJ19154011TA), 01008406821196031119040021RXJ19154010TA (RXJ19154010TA), 01008406821196031119040021RXJ19154007TA (RXJ19154007TA), 01008406821196031119050021RXJ19194009TA (RXJ19194009TA), 01008406821196031119040021RXJ19154003TA (RXJ19154003TA), Not Available (RXJ19224002TA), 01008406821196031119050021RXJ19224016TA (RXJ19224016TA), 01008406821196031119050021RXJ19204003TA (RXJ19204003TA), 01008406821196031119040021RXJ19154001TA (RXJ19154001TA), 01008406821196031119050021RXJ19194003TA (RXJ19194003TA), 01008406821196031119050021RXJ19224013TA (RXJ19224013TA), 01008406821196031119050021RXJ19194002TA (RXJ19194002TA), 01008406821196031119050021RXJ19194011TA (RXJ19194011TA), 01008406821196031119050021RXJ19204007TA (RXJ19204007TA), 01008406821196031119040021RXJ19154005TA (RXJ19154005TA), 01008406821196031119050021RXJ19204006TA (RXJ19204006TA), 01008406821196031119050021RXJ19194007TA (RXJ19194007TA), 01008406821196031119050021RXJ19194004TA (RXJ19194004TA), 01008406821196031119040021RXJ19154008TA (RXJ19154008TA), 01008406821196031119050021RXJ19204002TA (RXJ19204002TA), 01008406821196031119050021RXJ19224001TA (RXJ19224001TA), 01008406821196031119050021RXJ19194001TA (RXJ19194001TA), 01008406821196031119050021RXJ19224012TA (RXJ19224012TA), 01008406821196031119050021RXJ19194008TA (RXJ19194008TA), 01008406821196031119050021RXJ19194005TA (RXJ19194005TA), 01008406821196031119040021RXJ19154004TA (RXJ19154004TA), 01008406821196031119050021RXJ19194006TA (RXJ19194006TA), 01008406821196031119050021RXJ19224009TA (RXJ19224009TA), 01008406821196031119050021RXJ19204005TA (RXJ19204005TA), 01008406821196031119050021RXJ19194010TA (RXJ19194010TA), 01008406821196031119040021RXJ19154002TA (RXJ19154002TA)
- Recall started
- Sept. 16, 2019
- Device type
- Computer, Diagnostic, Programmable
- Root cause (FDA)
- Device Design
- What the company told customers
- GE Healthcare notified customers of the recall on about 09/16/2019, via "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were informed that certain CLab II Plus Amplifiers used with CardioLab/ComboLab systems have failed the patient leakage current test. There is a potential that if another device with electrical connectivity (e.g., ablation device, ECG monitor) fails, then the CLab II Plus Amplifier will not prevent an electric current from completing a circuit, which could cause an electrical shock to a patient. Instructions included that customers could continue to use the system by following the below instructions: 1. Test all equipment used in conjunction with CardioLab/ComboLab per manufacturer instructions on electrical safety to ensure that all equipment is in good operating condition and performs per manufacturer specifications. Note: Due to the potential safety risks of other devices used in conjunction with CLab II Plus Amplifiers in Electrophysiology Procedures, all equipment that can electrically contact the patient during a procedure should be checked against each devices safety conformances in manufacturer specifications. 2. If you suspect another connected device is or may be malfunctioning, discontinue use of the CLab II Plus Amplifier until all nonconforming equipment is corrected. If nonconforming equipment is used in conjunction with the CLab II Plus Amplifier, the CLab II Plus Amplifier will not provide necessary mitigation to prevent patient shock. Customers were also informed that GE Healthcare will correct all affected products and a GE Healthcare representative will contact customers to arrange for the correction. They were asked to complete and return the "MEDICAL DEVICE NOTIFICATION ACKNOWLEDGEMENT" form.
- Recall number
- Z-0257-2020
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-0257-2020. The agency record governs; check it and the company before acting.