GE Healthcare Voluson E8 ultrasound scanner with Software version 6.2.0 and 6.2.1 and DICOM stations with…

Nov. 7, 2007

Date posted by the FDA
Nov. 7, 2007
Status
Terminated
Filed by
Ge Healthcare
Product
GE Healthcare Voluson E8 ultrasound scanner with Software version 6.2.0 and 6.2.1 and DICOM stations with these software versions; GE Healthcare 9900 West Innovation Drive, Wauwatosa, WI 53226, USA.
Reason
When a spectrum of a Pulsed Wave Doppler shows an angle correction unequal to 0 is reloaded from the archive or the image is sent to a DICOM station, this angle correction is mistakenly disregarded resulting in an underestimation of the flow velocities.
Distribution pattern
Nationwide.
Quantity
521 Units
Codes and lots
Serial Numbers: D00210, D00223, D00225, D00239, D00241, D00293, D00326, D00347, D00415, D00425, D00435, D00445, D00455, D00456, D00459, D00460, D00465, D00467, D00468, D00469, D00480, D00481, D00484, D00485, D00487, D00490, D00491, D00493, D00495, D00502, D00503, D00505, D00512, D00516, D00517, D00523, D00525, D00531, D00532, D00537, D00539, D00541, D00542, D00543, D00546, D00548, D00550, D00551, D00552, D00554, D00558, D00560, D00561, D00562, D00563, D00564, D00565, D00566, D00567, D00568, D00569, D00570, D00571, D00572, D00573, D00574, D00575, D00576, D00577, D00579, D00580, D00581, D00582, D00625, D00627, D00628, D00632, D00633, D00636, D00638, D00639, D00640, D00649, D00652, D00653, D00655, D00665, D00666, D00668, D00108, D00123, D00134, D00139, D00168, D00173, D00176, D00178, D00182, D00185, D00186, D00187, D00189, D00192, D00193, D00194, D00195, D00198, D00199, D00200, D00203, D00209, D00214, D00220, D00224, D00226, D00227, D00228, D00230, D00231, D00232, D00233, D00236, D00242, D00243, D00244, D00246, D00249, D00252, D00254, D00258, D00282, D00291, D00292, D00295, D00296, D00297, D00299, D00300, D00302, D00303, D00304, D00305, D00306, D00307, D00310, D00312, D00314, D00316, D00317, D00318, D00319, D00327, D00330, D00335, D00336, D00337, D00339, D00340, D00343, D00344, D00346, D00349, D00350, D00351, D00352, D00354, D00355, D00356, D00359, D00361, D00362, D00364, D00366, D00367, D00372, D00374, D00375, D00388, D00391, D00392, D00393, D00395, D00396, D00397, D00398, D00399, D00402, D00403, D00404, D00405, D00406, D00411, D00413, D00418, D00420, D00421, D00422, D00423, D00426, D00451, D00180, D00379, D00387, D00430, D00433, D00444, D00454, D00464, D00470, D00476, D00497, D00509, D00513, D00522, D00526, D00540, D00544, D00545, D00547, D00549, D00553, D00555, D00556, D00557, D00559, D00585, D00586, D00103, D00155, D00166, D00250, D00251, D00255, D00325, D00331, D00595, D00584, D00280, D00342, D00369, D00148, D00213, D00301, D00428, D00601, D00602, D00603, D00604, D00324, D00151, D00471, D00519, D00593, D00466, D00494, D00659, D00432, D00473, D00321, D00591, D00599, D00153, D00217, D00222, D00248, D00437, D00514, D00322, D00449, D00474, D00475, D00478, D00504, D00507, D00508, D00521, D00528, D00529, D00533, D00534, D00583, D00634, D00635, D00641, D00644, D00645, D00663, D00126, D00128, D00240, D00272, D00277, D00279, D00332, D00341, D00407, D00419, D00486, D00424, D00499, D00288, D00308, D00394, D00483, D00143, D00117, D00138, D00150, D00165, D00167, D00183, D00218, D00245, D00253, D00259, D00261, D00264, D00315, D00328, D00378, D00383, D00385, D00431, D00462, D00613, D00618, D00121, D00141, D00154, D00170, D00171, D00190, D00234, D00256, D00267, D00311, D00313, D00461, D00145, D00146, D00196, D00201, D00205, D00206, D00207, D00257, D00263, D00266, D00268, D00410, D00438, D00439, D00458, D00479, D00488, D00489, D00496, D00511, D00518, D00520, D00527, D00538, D00594, D00596, D00598, D00600, D00605, D00621, D00660, D00661, D00179, D00202, D00204…
Recall started
July 6, 2007
Device type
System, Imaging, Pulsed Echo, Ultrasonic
Root cause (FDA)
Software design
What the company told customers
GE Healthcare sent customer letters on July 15, 2007 to affected customers. The letter included a short term mitigation recommendation and a long-term correction. The long-term correction will be software upgrade provided free of charge to all affected customers.
Recall number
Z-0113-2008

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-0113-2008. The agency record governs; check it and the company before acting.