FDA · Medical device recall
GE Healthcare Voluson E8 ultrasound scanner with Software version 6.2.0 and 6.2.1 and DICOM stations with…
- Date posted by the FDA
- Nov. 7, 2007
- Status
- Terminated
- Filed by
- Ge Healthcare
- Product
- GE Healthcare Voluson E8 ultrasound scanner with Software version 6.2.0 and 6.2.1 and DICOM stations with these software versions; GE Healthcare 9900 West Innovation Drive, Wauwatosa, WI 53226, USA.
- Reason
- When a spectrum of a Pulsed Wave Doppler shows an angle correction unequal to 0 is reloaded from the archive or the image is sent to a DICOM station, this angle correction is mistakenly disregarded resulting in an underestimation of the flow velocities.
- Distribution pattern
- Nationwide.
- Quantity
- 521 Units
- Codes and lots
- Serial Numbers: D00210, D00223, D00225, D00239, D00241, D00293, D00326, D00347, D00415, D00425, D00435, D00445, D00455, D00456, D00459, D00460, D00465, D00467, D00468, D00469, D00480, D00481, D00484, D00485, D00487, D00490, D00491, D00493, D00495, D00502, D00503, D00505, D00512, D00516, D00517, D00523, D00525, D00531, D00532, D00537, D00539, D00541, D00542, D00543, D00546, D00548, D00550, D00551, D00552, D00554, D00558, D00560, D00561, D00562, D00563, D00564, D00565, D00566, D00567, D00568, D00569, D00570, D00571, D00572, D00573, D00574, D00575, D00576, D00577, D00579, D00580, D00581, D00582, D00625, D00627, D00628, D00632, D00633, D00636, D00638, D00639, D00640, D00649, D00652, D00653, D00655, D00665, D00666, D00668, D00108, D00123, D00134, D00139, D00168, D00173, D00176, D00178, D00182, D00185, D00186, D00187, D00189, D00192, D00193, D00194, D00195, D00198, D00199, D00200, D00203, D00209, D00214, D00220, D00224, D00226, D00227, D00228, D00230, D00231, D00232, D00233, D00236, D00242, D00243, D00244, D00246, D00249, D00252, D00254, D00258, D00282, D00291, D00292, D00295, D00296, D00297, D00299, D00300, D00302, D00303, D00304, D00305, D00306, D00307, D00310, D00312, D00314, D00316, D00317, D00318, D00319, D00327, D00330, D00335, D00336, D00337, D00339, D00340, D00343, D00344, D00346, D00349, D00350, D00351, D00352, D00354, D00355, D00356, D00359, D00361, D00362, D00364, D00366, D00367, D00372, D00374, D00375, D00388, D00391, D00392, D00393, D00395, D00396, D00397, D00398, D00399, D00402, D00403, D00404, D00405, D00406, D00411, D00413, D00418, D00420, D00421, D00422, D00423, D00426, D00451, D00180, D00379, D00387, D00430, D00433, D00444, D00454, D00464, D00470, D00476, D00497, D00509, D00513, D00522, D00526, D00540, D00544, D00545, D00547, D00549, D00553, D00555, D00556, D00557, D00559, D00585, D00586, D00103, D00155, D00166, D00250, D00251, D00255, D00325, D00331, D00595, D00584, D00280, D00342, D00369, D00148, D00213, D00301, D00428, D00601, D00602, D00603, D00604, D00324, D00151, D00471, D00519, D00593, D00466, D00494, D00659, D00432, D00473, D00321, D00591, D00599, D00153, D00217, D00222, D00248, D00437, D00514, D00322, D00449, D00474, D00475, D00478, D00504, D00507, D00508, D00521, D00528, D00529, D00533, D00534, D00583, D00634, D00635, D00641, D00644, D00645, D00663, D00126, D00128, D00240, D00272, D00277, D00279, D00332, D00341, D00407, D00419, D00486, D00424, D00499, D00288, D00308, D00394, D00483, D00143, D00117, D00138, D00150, D00165, D00167, D00183, D00218, D00245, D00253, D00259, D00261, D00264, D00315, D00328, D00378, D00383, D00385, D00431, D00462, D00613, D00618, D00121, D00141, D00154, D00170, D00171, D00190, D00234, D00256, D00267, D00311, D00313, D00461, D00145, D00146, D00196, D00201, D00205, D00206, D00207, D00257, D00263, D00266, D00268, D00410, D00438, D00439, D00458, D00479, D00488, D00489, D00496, D00511, D00518, D00520, D00527, D00538, D00594, D00596, D00598, D00600, D00605, D00621, D00660, D00661, D00179, D00202, D00204…
- Recall started
- July 6, 2007
- Device type
- System, Imaging, Pulsed Echo, Ultrasonic
- Root cause (FDA)
- Software design
- What the company told customers
- GE Healthcare sent customer letters on July 15, 2007 to affected customers. The letter included a short term mitigation recommendation and a long-term correction. The long-term correction will be software upgrade provided free of charge to all affected customers.
- Recall number
- Z-0113-2008
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-0113-2008. The agency record governs; check it and the company before acting.