FDA · Medical device recall

AIR Technology Anterior Array…

Oct. 16, 2019 Class II

Date
Oct. 16, 2019
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
GE Healthcare, LLC
Product
AIR Technology Anterior Array, Catalogue Numbers: (a) 5746694, (b) 5746695, (c) 5746696, a receive-only RF coil designed for use with GE 3.0T MRI systems.
Reason
The outer seam of the coil, closest to the system cable, may separate and expose the porous material internal to the coil. This could result in user and/or patient contact with infectious agents since the internal porous material cannot be disinfected.
Distribution pattern
Worldwide Distribution - US Nationwide in the states of AZ, CA, MS, NY, WI, DE, and countries of France, Germany, Japan, Netherlands, Spain, Sweden,
Quantity
22 units
Codes and lots
Catalogue # UDI # Mfg. Lot or Serial # 5746694 0100840682138611111812132100339 339 5746694 0100840682138611111809122100324 324 5746694 0100840682138611111805312100311 311 5746694 0100840682138611111901022100340 340 5746694 0100840682138611111811302100334 334 5746694 0100840682138611111808302100323 323 5746694 0100840682138611111809132100325 325 5746694 0100840682138611111809192100326 326 5746694 0100840682138611111801292100303 303 5746694 0100840682138611111801232100301 301 5746694 0100840682138611111808152100319 319 5746694 0100840682138611111808272100320 320 5746694 0100840682138611111812042100335 335 5746694 0100840682138611111806072100313 313 5746694 0100840682138611111807182100315 315 5746694 0100840682138611111802122100307 307 5746694 0100840682138611111811092100328 328 5746694 0100840682138611111812072100337 337 5746694 0100840682138611111807232100316 316 5746695 0100840682138611111808282100321 321 5746695 0100840682138611111808292100322 322 5746696 0100840682138611111812132100338 338
Recall started
March 19, 2019
Device type
Coil, Magnetic Resonance, Specialty
Root cause (FDA)
Device Design
What the company told customers
The firm hand delivered the notices beginning on 03/19/2019. The notices described the separation of the outer seam and requested that the consignee check each coil component for tears or damage, and to discontinue use of product if product is suspected of not being in good condition. The affected products will be replaced. For questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative
Recall number
Z-0062-2020

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-0062-2020. The agency record governs; check it and the company before acting.